ChipFind - документация

Электронный компонент: C8147-38

Скачать:  PDF   ZIP
Photonic Multichannel Spectral
Analyzer Model: PMA-11
The PMA-11 is a spectral analyzer that integrates a
spectrograph and high performance multi-channel
photodetector in a single compact chassis. Light
collection is simplified through the use of optical fiber.
The diffraction grating of the spectrograph and multi-
channel photodetector are rigidly fixed, resulting in
excellent wavelength reproducibility. The wavelength
axis and spectral response characteristics are cali-
brated at the factory, so that spectral measurements
can be carried out easily and accurately.
The PMA-11 series offers four different multi-channel
photo-detectors to choose from, for additional flexibility
in grating selection, allowing the user to optimize the
performance for the application at hand.
Equipped with a standard SCSI interface, the PMA-11
is easily connected to any type of computer for data
collection and analysis.
q
Compact Integration of
a Spectrograph and Multi-
channel Photodetector
q
Easy Measurement Using
Optical Fiber Input
q
High Sensitivity
A compact unit containing a multi-channel photo-detector,and
power supply all in one.
Optical fiber input makes spectral measurements easier than ever.
FEATURES
q
Measurements of the spectrum are easier and more
accurate than ever before
The spectrum can now be easily measured by light collection through an
optical fiber. The wavelength axis and spectral response characteristics are
calibrated at the factory, so that spectral measurements can be carried out
easily and accurately.
q
Superb cost perfomance model : C5965
The C5965 uses a MOS linear image sensor realises high performance and
low cost.
q
High sensitivity model : C5966
The C5699 uses the CCD leaner image sensor has sensitivity a hundred
times better than the C5965 model.
q
Ultra-high sensitivity model : C7473-36
The C7473-36 consists the thermoelectric-cooling type BT-CCD image
sensors, which have a high quantum efficiency and a compact Czerny-
Turner type spectrograph. The simultaneous measurement of the wavelength
from an ultraviolet to a near-infrared region with high wavelength resolution
and high sensitivity is realised.
q
Near infrared model : C8147-34, C8147-38
The C8147 realises a simultaneous and high-resolution measurement of
absorption or reflection spectra in a near infrared wavelength region with
a wide dynamic range and a low noise.
q
High efficiency optics
Adoption of a
1mm bundle fiber and a bright spectrograph detects a
measured light efficiently.
q
Compact design
High performance is built in a small case. This completely new design
ensures that the PMA-11 will fit anywhere.
q
External synchronisation can be used
Measurements can now be carried out synchronised to external trigger
signals, allowing measurement of pulse phenomena.
q
Standard SCSI interface allows connection to computer
APPLICATIONS
q
UV to visible spectroscopy
q
Fluorescence spectroscopy
q
Raman scattering
q
Chemiluminescence
analysis
q
Liquid chromatography
Personal computer
POWER
SIGNAL INPUT
C7473
HAMAMATSU
Basic software
Microscope
SCSI
Main unit
Fiber input optics
PMA-11 Standard Configuration
C5965
(MOS linear image sensor model)
C5966
(CCD image sensor model)
C7473
(BT- CCD image sensor model)
C8147
(InGaAs linear image sensor model)
Light source
Microscope
attachment
Main unit
Fiber input optics
Basic software
[Standard Configuration]
* SCSI cable not included
q
Gas chromatography
q
ICP emission analysis
q
Discharge emission analysis
q
Combustion analysis
q
Micro spectroscopy
q
Water quality testing
q
Evaluation of light sources
q
Chromaticity
measurements
q
Impurities testing
q
Thin film thickness monitors
q
UV-ray monitors
q
Plasma monitors
q
Fruit tester
q
Plastic sorting
q
Color filter testing
[ Industrial applications ]
[ Scientific applications ]
1
1
1
1
Luminescence spectrum of a deuterium lamp
MEASUREMENT EXAMPLES
3
Luminescence spectrum of an LED
0.
5
Fluorescence spectrum of fluorosein
2
Absorption spectrum of a didymium filter
4
Chromaticity coordinates of an LED
6
3-d display of plasma emission spectra
7
Transmittance spectra in near infrared reagion
Dotted line: Compact disc
Solid line: PET botle
8
Transmittance spectra in near infrared reagion
Dotted line : Styren monomer
Solid line : Polystyren
300
450
600
750
900
Wavelength (nm)
Relativ
e Intensity
500
900
800
700
600
Wavelength (nm)
Optical Density
0.0
0.5
1.0
1.5
2.0
300
450
600
750
900
Wavelength (nm)
300
450
600
750
900
Wavelength (nm)
900
1050
1200
1350
1500
Wavelength (nm)
1650
900
1050
1200
1350
1500
Wavelength (nm)
1650
Relativ
e Intensity
Relativ
e Intensity
T
r
ansmittance (%)
0
0
0
60
70
80
90
0
25
50
75
100
T
r
ansmittance (%)
0.00
0.15
0.30
0.45
0.60
[ x ]
0.0
0.2
0.4
0.6
0.8
[ y ]
Angle of field
: 2
Chromaticity coordinates
Dominant wavelength
Excitation purity
Tristimulus value
X : 58.25
Y : 100.00
Z : 0.96
x : 0.3659
y : 0.6281
: 559.04 nm
: 98.97%
Type No.
C5965-31
C5966-3x
C7473-36
Photodetector
MOS linear image sensor
CCD linear image sensor
Cooling temepareture
non-cooling
0C
Read-out noise
10 000 electrons
60 electrons
10 electrons
Dark current
12,500 electrons/scan ( at 25C; 20ms )
512 electrons/scan( at 0C; 20ms )
75 electrons/scan ( at -15C; 20ms )
C8147-34
InGaAs linear image sensor
No. of photosensitive device channels
1024 ch
1024 ch
1024 ch
256 ch
Channel size
25
m(H)
2.5 mm (V)
24
m (H)
3.07 mm(V)
24
m(H)
2.928 mm(V)
50
m(H)
250
m (V)
0C
12,500 electrons
20,000 electrons/scan ( at 0C; 5ms )
A/D resolution
Spectrograph F number
Spectrograph type
Simultaneous measurement
wavelength range
Wavelength resolution
*
300 nm to 800nm
< 3 nm(FWHM)
200 nm to 950 nm
< 2 nm(FWHM)
900 nm to 1650 nm
< 9 nm(FWHM)
x=1 300 nm to 800 nm
x=2 200 nm to 400 nm
x=3 600 nm to 1000nm
x=1 < 3 nm(FWHM)
x=2 < 1.5 nm(FWHM)
x=3 < 2.5 nm(FWHM)
Effective Light-receiving area of optical fiber
Interface
SCSI
Line voltage
AC100V to 240V
10%, 50, 60Hz
1mm
Optical fiber length
1.5m
Exposure time
5 ms to 32 s
20ms to 32s
External trigger input
TTL level / High impedance
16bit
3
4
Concave spherical grating type
Czerny-Turner type
-15C
BT- CCD linear image sensor
SPECIFICATIONS
*
Tested by the bright-line spectrum of Hg-Ar lamp (at 312.57nm, 435.84nm, 546.07nm, 696.54nm, 1013.98nm)
DIMENSIONAL OUTLINES (Unit :mm)
q
C5965. C5966 (approx. 4.5kg)
6
1
40
1
260
1
3
0.5
98.5
1
170
1
232
1
333
1
POWER
SIGNAL INPUT
PMA-11
PHOTONIC MULTI-CHANNEL ANALYZER
q
C7473 (approx. 5kg)
6
1
25
1
325
1
3
0.5
98.5
1
200
1
262
1
383
1
POWER
SIGNAL INPUT
C7473
PHOTONIC MULTI-CHANNEL ANALYZER
6
1
25
1
375
1
3
0.5
123.5
1
200
1
262
1
433
1
POWER
SIGNAL INPUT
PMA-11
PHOTONIC MULTI-CHANNEL ANALYZER
q
C8147 (approx. 7.5kg)
12-0.1
+0
80
1500
50
10-0.1
+0
Bundle diameter 1.0
3
0.2
C1
q
Fiber input optics (approx.100g)
Cat. No. SSCS1040E07
JUL /2002 HPK
Created in Japan (PDF)
PVC sheathed SUS
flexible tube ( 7)
15 strands
q
Basic software
Spectral measurement
Measurement
Reflection spectra measurement
functions
Absorption spectra measurement
Color measurement
Temporal resolution
Temporal fluctuation of spectra over time
measurement functions
Temporal fluctuation of reflection factor and
transmittance over time
Data acquisition
Exposure time
condition setting
Memory integration count times
Temporal fluctuation measurement
Calibration and
Wavelength axis
correction
Sensitivity uniformity
Dark current
Display functions
Spectrum (non-limited accumulation)
Temporal fluctuation of waveform over time
(non-limited accumulation)
Chromaticity diagram
Wavelength axis display
Wavelength (nm) ,Wavenumber (cm
-1
),
Energy (eV)
Brightness axis display
Linear, logarithm
Wavelength (Wavenumber etc.) vs, intensity
Cursor analysis
Peak detection
functions
FWHM between two cursors
Integrated intensity
Other analytical
Smoothing
functions
Differential waveform
Color measurement
q
Main unit
Homepage Address http://www.hamamatsu.com
5
Product and software package names noted in this documentation are trademarks or registered trademarks of their respective manufacturers.
q
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
.
q
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions.
Specifications and external appearance are subject to change without notice..
2001Hamamatsu Photonics K.K.
ISO 9001
Certificate: 09 105 79045
HAMAMATSU PHOTONICS K.K., Systems Division
812 Joko-cho, Hamamatsu City, 431-3196, Japan, Telephone: (81)53-431-0124, Fax: (81)53-435-1574, E-mail:export@sys.hpk.co.jp
U.S.A. and Canada: Hamamatsu Photonic Systems: 360 Foothill Road, Bridgewater, N.J. 08807-0910, U.S.A., Telephone: (1)908-231-1116, Fax: (1)908-231-0852, E-mail: usa@hamamatsu.com
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658, E-mail: info@hamamatsu.de
,
France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10, E-mail: infos@hamamatsu.fr
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire, AL7 1BW, U.K., Telephone: (44) 1707-294888, Fax: (44) 1707-325777, E-mail: info@hamamatsu.co.uk
North Europe: Hamamatsu Photonics Norden AB: Smidesvgen 12, SE-171-41 Solna, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01, E-mail: info@hamamatsu.se
-
Italy: Hamamatsu Photonics Italia S.R.L.: Strada della Mois, 1/E 20020 Arese (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741, E-mail: info@hamamatsu.it
C8147-38
-10C
2.5
10
7
electrons/scan ( at -10C; 5ms )
1600 nm to 2350 nm
< 9 nm(FWHM)
5 ms to 50 ms (typ.)