ChipFind - документация

Электронный компонент: L7810

Скачать:  PDF   ZIP
NEW
Calibrated Xenon Light Source
L7810
FEATURES
s
High Stability
q
Light Intensity Drift (Typ.)
q
Comparison of Spectral Irradiance (Reference data)
: 50 cm away from light source
Stability is one of the important characteristics of calibrated
light sources. Hamamatsu's excellent cathode technology
and optimized lamp housing realize high stability and good
repeatability.
s
High Power
A Xenon lamp emits high intense UV, visible and IR light so
that a reliable measurement can be possible even in the
range of 300 nm to 400 nm, where either a deuterium lamp or
a tungsten lamp doesn't deliver enough light.
is a logo mark for an accreditation
system based on Japanese Meas-
urement Law. Our Planning & Engi-
neering Division is accredited in
accordance with ISO/IEC Guide 25:
1990 in the field of photometry cali-
bration.
(Our accreditation number is 0081.)
[JCSS: Japan Calibration Service
System (The Japanese Goverment
accreditation scheme for calibration
laboratories and a signatory of
APLAC Mutual Recognition Arrange-
ment)]
v
LEFT: LAMP HOUSING, RIGHT: POWER SUPPLY
OPERATING TIME (h)
RELATIVE INTENSITY (%)
10
20
0
50
40
30
20
10
80
70
60
110
100
90
30
40
50
60
70
80
90
100
110
CONVENTIONAL HALOGEN LAMP
CONVENTIONAL DEUTERIUM LAMP
CALIBRATED XENON LIGHT SOURCE
200
250
300
350
400
450
500
550
600
650
700
750
800
0.001
0.01
0.1
1
10
100
WAVELENGTH (nm)
SPECTRAL IRRADIANCE (
Wcm
-2
nm
-1
)
TLSOB0065EA
TLSOB0063EA
TLSO1039E01
JUN. 2000 SI
Printed in Japan(1000)
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein. 2000 Hamamatsu Photonics K.K
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
SPECIFICATIONS
DIMENSIONAL OUTLINES
(Unit: mm)
HAMAMATSU PHOTONICS K.K., Electron Tube Center
314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de
France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: france@hamamatsu.com
United Kingdom: Hamamatsu Photonics UK Limited: Lough Point, 2 Gladbeck Way, Windmill Hill, Enfield, Middlesex EN2 7JA, United Kingdom, Telephone: 44(20)8-367-3560, Fax: 44(20)8-367-6384 E-mail: info@hamamatsu.co.uk
North Europe: Hamamatsu Photonics Norden AB: Smidesvgen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se
Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it
HOMEPAGE URL http://www.hamamatsu.com
Lamp
Input Wattage
Spectral Distribution
Calibrated Range
A
Type of Calibration
Calibration Unit
Stability
Window Material
(Lamp Housing)
Warm-up Time
Recommended Calibration
Period
B
Guaranteed Life
C
Fluctuation (p-p)
Drift
NOTE:
A
The calibrated value is the light power at 50 cm away from front surface of lamp housing.
B
This is the recommended calibration period when high reliability is required. Please contact to our sales office for re-calibration fee.
C
This is the guaranteed time the unit satisfies requirements as calibrated light source.
D
This is an accessory coming with the unit.
Xenon Lamp
150 W
185 nm to 2000 nm
250 nm to 800 nm
Spectral Irradiance
W.cm
-2
.nm
-1
1.0 % Max.
0.5 %/h Max.
Synthetic Silica
10 min
100 hours or 1 year whichever
comes first
500 hours
Input Voltage
(switched manually)
Power Consumption
Cooling (Lamp Housing,
Power Supply)
Ambient Temperature
Humidity
Dimensions
(W
H
D)
Weight
Lamp Housing
Power Supply
Case
D
Lamp Housing
Power Supply
Case
D
100 Vac/115 Vac/200 Vac/230 Vac
(10 %), 50/60 Hz
500 VA
Forces Air Cooling
(cooling fan is included)
0
C to 40
C
30 % to 90 %
100 mm
280 mm
100 mm
144 mm
176 mm
280 mm
595 mm
425 mm
345 mm
Approx. 3.5 kg
Approx. 10 kg
Approx. 7.8 kg
100
COOLING FAN
M37 P=1
OUTPUT
WINDOW
( 28)
50
140
280
5
60
77
5
100
5
78
39
78
39
[FRONT VIEW]
[BOTTOM VIEW]
[SIDE VIEW]
CONNECTOR *
SIDE COVER
MOUNTING
SCREW
4-M3
HEX SOCKET
HEAD BOLT
SIDE COVER
(REMOVE
THIS COVER
WHEN
REPLACING
LAMP)
M4
4-M4
PLASTIC FEET ARE ATTACHED TO
THE BOTTOM OF THE UNIT BUT
CAN BE REMOVED TO USE THEIR
HOLES FOR INSTALLATION OF THE
LAMP HOUSING.
REFERENCE SURFACE
OF CALIBRATION
* For cable to connect lamp housing with power supply, length: 2000
CURRENT SEL
M 1A/125V
TIME COUNTER
0.0
XENON LAMP POWER SUPPLY
hours
LAMP ON
7.5
8.0
8.5
POWER
[Adc]
POWER
INTER LOCK
OVER HEAT
[FRONT VIEW]
[SIDE VIEW]
[REAR VIEW]
OUTPUT
SN.
VOL. SEL
100/115/200/230
50/60 HZ
144
176
280
CONNECTOR *
AC INPUT CONNECTOR
TLSOA0066EA
TLSOA0067EA
LAMP HOUSING