SPECIFICATION
Parameter
C7860
Input
Output
Ambient Temperature
Cooling
Dimensions ( W
H
D )
Weight
Input Voltage
Input Current
Output Voltage
Output Current
Current Fluctuation (p-p)
Current Drift at + 25
C
Heater Voltage
Heater Current
Warm-up Time
Trigger Voltage
With Load
Without Load
Output Voltage
Output Current
Operation
Storage
2
80
160
300
15
0.5
0.1
See below
See below
25
5
600
50
0 to +40
-40 to +60
V
A Max.
V dc Typ.
V dc Min.
mA
% Max.
%/h Max.
--
--
s
V peak
C
C
--
mm
kg
20 CFM of Forced Air Cooling
100
118
36.2
0.17
24
2.4 (dc)
M7628
Unit
--
113
122
220
2.7
90 to 115 / 180 to 250 (ac)
( Automatic)
DEUTERIUM LAMP
RELATED PRODUCTS
Information furnished by HA MAM ATS U is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein.
2000 Hamamatsu Photonics K.K.
Subject to local technical requirements and regulations, availability of products included in this promotional material may var y. Please consult with our sales office.
HEATER VOLTAGE AND CURRENT
Warm-up
Voltage
(V dc)
Current
(A dc)
Operation
Applicable Lamps
Type Number
C7860/M7628-2510
C7860/M7628-2517
A
C7860/M7628-3000
A
C7860/M7628-1035
A
C7860/M7628-1070
C7860/M7628-1555
A
NOTE:
A
C7860 series are manufactured only when the order is placed.
2.5
0.15
2.5
0.15
3
0.15
10
0.5
10
0.5
15
0.75
4
4
5
0.8
1.2
0.5
Voltage
(V dc)
Current
(A dc)
1
0.05
1.7
0.1
0
3.5
0.2
7
0.35
5.5
0.3
1.8
3.3
0
0.3
1
0.3
L6565, L7293, L7303, L6999, L7307, L7174
L7297, L7298, L6303, L6304, L7304, L7306
L6566, L6305, L6306
L7291, L7294, L6307, L6308, L7292
L7296, L7295, L6309, L6310
L6311, L6312
FEATURES
s
Compact
s
High Cost Performance
s
Designed based on EN/UL standard (M7628)
The C7860 designed based on EN standard only.
s
Constant Current Power Supply
POWER SUPPLY C7860/OEM POWER SUPPLY M7628
Left: C7860, Right: M7628
DEUTERIUM LAMP POWER SUPPLY C7860/M7628 SERIES
TLSOA0057EA
TLSOA0061EA
s
Dimensional Outline (Unit: mm)
1. High voltage is used to operate the lamp. Use extreme caution to prevent electric shocks.
2. The guarantee period will be one year after shipment. The area of guarantee is limited to replacement of the faulty lamp. Faults resulting
from natural disasters and incorrect usage will also be excluded from guarantee.
Precautions When Using Deuterium Lamps
102
7.7
118
36.2 MAX.
1.5
6.2
86.4
7
100
28.5
9.5
8.1
#6-32 SCREW TERMINALS, 8 PLACES
4-M3
MAXIMUM
COMPONENT
HEIGHT
POSITION
DEUTERIUM
LAMP
(1) ANODE
(2) APERTURE
(3) HEATER
(4) GROUND (GND)
(5) +24 VDC 10 %
(6) +24 RETURN (GND)
(7) +5 VDC (+)
(8) +5 RETURN (-)
DISABLE INPUT CIRCUIT
DISABLE FUNCTION
DISABLE = +5 V BETWEEN PINS 7 & 8 POLARITY AS INDICATED, LOAD = 2 mA
ENABLE = 0 V BETWEEN PINS 7 & 8 OR PINS 7 & 8 OPEN
OPTO COUPLER
7
1.8 k
8
INPUT
POWER
DISABLE
These product conforms to SAFETY/ EN61010-1 of the European Union.
The M7628 is recognized upon UL 1950 (File No.191652).
200
113
122
FRONT VIEW
SIDE VIEW
TOP VIEW
REAR VIEW
C7860
M7628
SPECIFICATION
NOTE:
A
A8087 cable should be necessary for connection between C7860 series and E8039.
APPLICATIONS
s
Light sources in various experiments
s
Light sources in spectrophotometry
FEATURES
s
Compact
s
High Stability
s
High Safety, Interlock Function
s
No Optical Alignment Required
s
Easy Replacement of Optional Parts
DEUTERIUM LAMP HOUSING E8039
Parameter
Suitable Lamp
Lamp Power Consumption
Spectral Distribution
Anode Current
Tube Drop Voltage
Material
Dimension
Optical Axis Height
Optional Parts SMA Fiber Guide Adaptor with Lens (for thin diameter fiber)
Collimating Lens (Material: Fused Silica, Dimension: 10 mm, Focal Point: 20 mm, Light output: collimating)
Power Supply C7860 series, Connection cable A8087 and Lamp L6999-51 are optional.
L2D2 L6999-51
Approx. 30
185 to 400
80
300
Fused Silica
10
72 (include feet height)
Forced Air Cooling (Built-in cooling fan)
0 to +40
+5 to +35
30 to 90
120
107
140 (exclude projection part)
Approx. 2.3
C7860-2510
A
Lamp
Output Window
Cooling Function
Operating Ambient Temperature
Recommended Operating Temperature
Operating Humidity
Dimensions ( W
H
D )
Weight
Suitable Power Supply Type No.
Description/Value
--
W
nm
V dc Typ.
mA dc
--
mm
mm
--
C
C
%
mm
kg
--
Unit
This lamp housing was designed to allow
easy operation of deuterium lamps such as
L2D2 lamps and provide full lamp
performance. It accommodates a lamp with
a flange so that no optical alignment is
required. The built-in interlock and forced-air
cooling functions ensure high safety.
Collimating lenses and fiber guide adaptors
are also available as easy-to-replace
options, which easily attach to the light exit
and allow obtaining the desired light beam.
Compact case allows easy handling and
installation.
Lamp housing design allows the L2D2 lamp to fully
achieve its highly stable operation (fluctuation(p-p)
within 0.05%).
The interlock function turns off the lamp if the
cover comes off. The surface temperature of the
lamp housing is maintained below +50
C.
Designed to accommodate an L2D2 lamp with a
flange which eliminates arc point shift. Absolutely
no optical alignment is required over the life of the
lamp or when replacing the lamp.
As easy-to-replace options, collimating lenses and
fiber guide adaptors that easily attach to the light
exit are available, allowing you to obtain the
desired light beam.
DEUTERIUM LAMP HOUSING E8039
TLSO1036E03
OCT. 2000 IP
Printed in Japan (1000)
TLSOA0062EA
s
Dimensional Outlines (Unit: mm)
1. High voltage is used to operate the lamp. Use extreme caution to prevent electric shocks.
2. The guarantee period will be one year after shipment. The area of guarantee is limited to replacement of the faulty lamp. Faults resulting
from natural disasters and incorrect usage will also be excluded from guarantee.
Precautions When Using Deuterium Lamps
140
9
60
ARC POINT
OUTPUT
WINDOW
LAMP REPLACEMENT
PORT
63
107
9.5
120
10
HAMAMATSU PHOTONICS K.K., Electron Tube Center
314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de
France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: infos@hamamatsu.fr
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: info@hamamatsu.co.uk
North Europe: Hamamatsu Photonics Norden AB: Smidesvgen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se
Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it
HOMEPAGE URL http://www.hamamatsu.com