ChipFind - документация

Электронный компонент: 19C005PG1K

Скачать:  PDF   ZIP
Pressure Sensors
Low-Cost, Stainless Steel, Isolated Sensors
19 mm Series
Sensing and Control
FEATURES
Low cost
Rugged, isolated stainless
steel package
Small size
Reliable semiconductor
technology
Calibrated and
temperature compensated
Absolute and gage
pressures
Vacuum compatible,
isolated sensors
0 psi to 3 psi to, 0 psi to
500 psi
TYPICAL APPLICATIONS
Industrial controls
Process control systems
Industrial automation and
flow control
Pressure calibrators









WARNING
PERSONAL INJURY
DO NOT USE these
products as safety or
emergency stop devices or
in any other application
where failure of the product
could result in personal
injury.
Failure to comply with
these instructions could
result in death or serious
injury.
Honeywell's stainless steel 19C, 19U, and 19 Vacuum Gage Series
sensors were developed for pressure applications that involve measurement
of hostile media in harsh environments compatible with 316 stainless steel.
The special Vacuum Gage Series sensors are specifically designed for
applications that can be exposed to vacuum.
The 19C Series are calibrated and temperature compensated. They are
available for use with either a constant voltage or current source. They
feature a variety of pressure connections to allow use in a wide range of
OEM (Original Equipment Manufacturer) equipment. The 19U Series is
uncompensated for applications that use their own specialized circuit
designs.
The 19C and 19U Series sensors are rugged and reliable transducers for
use in a wide variety of pressure sensing applications where corrosive
liquids or gases are monitored.
19C, 19U, AND 19 VACUUM GAGE EQUIVALENT CIRCUIT


WARNING
MISUSE OF DOCUMENTATION
The information presented in this product sheet is for reference only. Do
not use this document as a product installation guide.
Complete installation, operation, and maintenance information is provided
in the instructions supplied with each product.
Failure to comply with these instructions could result in death or
serious injury.
Pressure Sensors
Low-Cost, Stainless Steel, Isolated Sensors
19 mm Series
2 Honeywell
Sensing and Control
PRESSURE SENSOR CHARACTERISTICS ALL DEVICES
Environmental specifications
Compensated temperature range
0 C to 82 C [32 F to 180 F]
Operating temperature range
-40 C to 125 C [-40 F to 257 F]
Storage temperature range
-40 C to 125 C [-40 F to 257 F]
Vibration
10 G at 20 Hz to 2000 Hz
Shock
100 G for 11 ms
Life
1 million cycles min.
Insulation resistance
100 m
at 50 Vdc
Maximum ratings
Voltage version "K"
Supply voltage V
S
= +15 Vdc
Voltage version "L"
Supply current I
S
= +2.0 mA

PRESSURE RANGE SPECIFICATIONS (19C, 19U)
Catalog Listing
Pressure range
Proof pressure
(7)
Burst
pressure
(8)
19(C,U) 003P G (1...9) (K/L)
0 psig to 3 psig
9 psig
15 psig
19(C,U) 005P G (1...9) (K/L)
0 psig to 5 psig
15 psig
25 psig
19(C,U) 010P G (1...9) (K/L)
0 psig to 10 psig
30 psig
50 psig
19(C,U) 015P (A,G,V) (1...9) (K/L)
0 psi to 15 psi
45 psi
75 psi
19(C,U) 030P (A,G,V) (1...9) (K/L)
0 psi to 30 psi
90 psi
150 psi
19(C,U) 050P (A,G,V) (1...9) (K/L)
0 psi to 50 psi
150 psi
250 psi
19(C,U) 100P (A,G,V) (1...9) (K/L)
0 psi to 100 psi
300 psi
500 psi
19(C,U) 200P (A,G,V) (1...9) (K/L)
0 psi to 200 psi
600 psi
1000 psi
19(C,U) 300P (A,G,V) (1...9) (K/L)
0 psi to 300 psi
900 psi
1500 psi
19(C,U) 500P (A,G,V) (1...9) (K/L)
0 psi to 500 psi
1200 psi
2400 psi
Notes:
7. The maximum pressure that can be applied without changing the transducer's performance or accuracy.
8.The maximum pressure that can be applied to a transducer without rupture of either the sensing element or transducer case.

19C SERIES PERFORMANCE CHARACTERISTICS
(1)
Characteristic Minimum
Typical
Maximum
Unit
Zero
pressure
offset
-2 0 2 mV
Full-scale span
(2)
98 100 102 mV
Full-scale span (3 psi and 5 psi "K" units only)
(2)
48 50 52 mV
Pressure non-linearity
(3)
0.1
0.25
%FSS
Pressure hysteresis
(3)
0.015 0.030 %FSS
Repeatability
0.010 0.030 %FSS
Temp. effect on span
(4)
0.5 1.0 %FSS
Temp. effect on offset
(4)
0.5 1.0 %FSS
Temp. effect on span (0 psi to 3 psi and 0 psi to 5 psi only)
(4)
2.0
%FSS
Temp. effect on offset (0 psi to 3 psi and 0 psi to 5 psi only)
(4)
2.0
%FSS
Thermal hysteresis (0 C to 82 C [32 F to 180 F])
0.1
0.3
%FSS
Long-term stability of offset and span
(5)
0.1 0.3 %FSS
Response time
(6)
0.1
ms
Common mode voltage (voltage version "K")
(9)
0.5 1.3 2.0 Vdc
Input
impedance
(current
version
"L")
2.0 4.5 8.0 k
Input impedance (voltage version "K")
8.0
25
50
k
Output
resistance
3.0 4.5 6.0 k


Pressure Sensors
Low-Cost, Stainless Steel, Isolated Sensors
19 mm Series
Honeywell
Sensing
and
Control
3
19U SERIES PERFORMANCE CHARACTERISTICS
(1)
Characteristic Minimum
Typical
Maximum
Unit
Zero pressure offset
-7.5
1.0
7.5
mV/V
Full-scale span
(2)
35 40 60 mV/V
Full-scale span (3 psi and 5 psi only)
(2)
20 30 40 mV/V
Pressure non-linearity
(3)
0.1
0.25
%FSS
Pressure hysteresis
(3)
0.015
0.030
%FSS
Repeatability
0.010
0.030
%FSS
Temp. coefficient of resistance
(10)
(0 C to 82 C [32 F to 180 F])
2700
3420
4500
ppm/C
Temp. coefficient of offset
(10)
(0 C to 82 C [32 F to 180 F])
30
V/V/C
Temp. coefficient of span
(10)
(0 C to 82 C [32 F to 180 F])
360
720
1260
ppm/C
Thermal hysteresis (0 C to 82 C [32 F to 180 F])
0.1
0.3
%FSS
Long-term stability of offset and span
(5)
0.1
0.3
%FSS
Response time
(6)
0.1
ms
Input resistance
4.00
4.75
6.0
k
Output resistance
4.00
4.75
6.0
k
19C VACUUM GAGE COMPENSATED SERIES PERFORMANCE CHARACTERISTICS
(1)
Characteristic Minimum
Typical
Maximum
Unit
Zero pressure offset (at 0 psig)
-2
0
2
mV
Full-scale span
(2)
(0 psig to positive full-scale)
98
100
102
mV
Pressure non-linearity
(3)
0.1
0.25
%FSS
Pressure hysteresis
(3)
0.015
0.030
%FSS
Repeatability
0.010
0.030
%FSS
Temp. effect on span
(4)
0.5
1.0
%FSS
Temp. effect on offset
(4)
0.5
1.0
%FSS
Thermal hysteresis (0 C to 82 C [32 F to 180 F])
0.1
0.3
%FSS
Long-term stability of offset and span
(5)
0.1
0.3
%FSS
Response time
(6)
0.1
ms
Common mode voltage (voltage version "K")
(9)
0.5
1.3
2.0
Vdc
Input impedance (current version "L")
1.5
4.5
8.0
k
Input impedance (voltage version "K")
8.0
25
50
k
Output resistance
3.0
4.5
6.0
k
19U VACUUM GAGE UNCOMPENSATED SERIES PERFORMANCE CHARACTERISTICS
(1)
Characteristic Minimum
Typical
Maximum
Unit
Zero pressure offset
-7.5
1.0
7.5
mV/V
Full-scale span
(2)
(0 psig to positive full-scale)
35
40
60
mV/V
Pressure non-linearity
(3)
0.1
0.25
%FSS
Pressure hysteresis
(3)
0.015
0.030
%FSS
Repeatability
0.010
0.030
%FSS
Temp. coefficient of resistance
(10)
(0 C to 82 C [32 F to 180 F])
2700
3420
4500
ppm/C
Temp. coefficient of offset
(10)
(0 C to 82 C [32 F to 180 F])
30
V/V/C
Temp. coefficient of span
(10)
(0 C to 82 C [32 F to 180 F])
360
720
1260
ppm/C
Thermal hysteresis (0 C to 82 C [32 F to 180 F])
0.1
0.3
%FSS
Long-term stability of offset and span
(5)
0.1
0.3
%FSS
Response time
(6)
0.1
ms
Input resistance
4.00
4.75
6.0
k
Output resistance
4.00
4.75
6.0
k
Pressure Sensors
Low-Cost, Stainless Steel, Isolated Sensors
19 mm Series
4 Honeywell
Sensing and Control
NOTES
1. Reference conditions (unless otherwise noted): T
A
= 25 C [77 F]; 19C Supply V
S
= 10 Vdc 0.01 Vdc or
I
S
= 1.5 mA 0.0015 mA; 19U Supply V
S
= 5 Vdc 0.01 Vdc or I = 1.0 mA 0.0015 mA
2. Full-scale span (FSS) is the algebraic difference between the output voltage at full-scale pressure and the output at zero
pressure. FSS is ratiometric to supply voltage.
3. Pressure non-linearity is based on best-fit straight line from zero to the full-scale pressure. Pressure hysteresis is the
maximum output difference at any point within the operation pressure range for increasing and decreasing pressure.
4. Maximum error band of the offset voltage or span over the compensated temperature range, relative to the 25 C [77 F]
reading.
5. Long-term stability over a six-month period.
6. Response time for a 0 psi to FSS pressure step change, 10 % to 90 % rise time.
9. Common mode voltage as measured from output to ground. For higher levels of common mode voltage, contact factory.
10. The error band resulting from a maximum deviation of a transducer's output parameter (offset, span, or resistance) as
temperature is varied 25 C [77 F] to any other temperature within the specified range (0 C to 82 C [32 F to 180 F]).
This parameter is not 100 % tested and is guaranteed by process design and tested on a sample basis only.
Temperature coefficient of span is evaluated using a constant current source.

PHYSICAL DIMENSIONS, for reference only, mm [in]

Package 1 (see note): Cell Package
Note: Non-concentricity effects at the diaphragm weld area may cause runout of up to 0.006 in between the upper and lower portions of
the sensor body. (It is recommended to use a counter bore to mate with this device to allow for this non-concentricity).

Package 2: Flush Mount Package
Pressure Sensors
Low-Cost, Stainless Steel, Isolated Sensors
19 mm Series
Honeywell
Sensing
and
Control
5
Package 3: Flush Mount with Flange
Package 4: Male 1/8-27 NPT
Package 5: Male 1/4-18 NPT